AUTOMATED INSTALLATION FOR DETERMINING THE WEAR OF THE CATALYST

Back to search results
AUTOMATED INSTALLATION FOR DETERMINING THE WEAR OF THE CATALYST
Manufacturer:
NPF "META-KHROM"
Location:
Yoshkar-Ola
Applying:
An automated installation for determining the wear of various catalysts, including KPS (FCC) for cracking with a fluidized bed by an air jet and is designed to determine the comparative wear characteristics of the KPS (FCC) of the catalyst by friction of an air jet according to ASTM D5757-11.
It is used for testing with irregularly shaped spherical sludge particles (size from 10 to 180 microns).
Abrasion is achieved by fluidization of the sample under the influence of jets of moist air. The percentage of fine particles (less than 20 microns in size) after a 5-hour test is an air jet abrasion index (AJI).
technical specifications
The range of operating air pressures for the power supply of the installation is 0.8±0.1 MPa
The range of air flow rates for the power supply of the installation is not less than (l/min) 20
Absolute error of the air flow rate maintained automatically at a flow rate of 10 l/min. no more, l/min. ± 0.05
Power consumption not more than 0.2 kW
Overall dimensions, mm and weight kg. 560x760x1700, 30
Abrasion pipe made of steel 12x18 n10t with dimensions, mm 710±10, 35 ± 0.5
Three sapphire nozzles with a diameter and length, mm 0.381±0.005, 2±1.0
Continuous operation mode
Control: microprocessor with tactile control for setting the display and automatic maintenance of installation parameters and mathematical processing of analysis results 60
The convergence of the results when testing the same catalyst is not more than 0.1%