Back to search results

THE DEVICE FOR ACHIEVING BALANCE

Back to search results
THE DEVICE FOR ACHIEVING BALANCE
When analyzing gases dissolved in transformer oil using the ARP method (equilibrium vapor analysis), mixing of the contents of the syringe with oil is used to accelerate the extraction of gases (to achieve an equilibrium distribution of gases dissolved in oil between the liquid and gas phases). For this purpose, as well as automation of the mixing process, a special autonomous device for achieving equilibrium is produced.
technical specifications
The rotation speed of the syringe drum is 30 rpm./min.
The rotation time (determined by the internal timer, after pressing the "Start" button) is 10 minutes.
The number of syringes placed is 4.
The volume of syringes is 20 or 50 ml.
Power consumption, no more than — 14 VA.
Overall dimensions, no more than 210×280×270 mm.
Weight, not more than — 5 kg.