Installation for determining the melt flow index of thermoplastics with auto-cutting PTR-LAB-11
Technical specifications:
Melt flow rate for 10 min, g 0.1 - 99.9
The dimensions of the capillary opening, mm
dia. 2,095
diam. 1,048 (optional)
diam. 1,180 (optional)
Set of loads, kg 0,325-5
10-21,6 (optional)
Temperature range in the working chamber, °C +50...+450
Accuracy of temperature maintenance, °C:
- in the range of 50...300 °C: ± 0.1
- in the range of 300...350 ° C: ± 0.2
- in the range of 350...450°C: ±0.3
The discreteness of setting the temperature of the chamber, ° C 1
Time to enter the operating mode, no more, min 40
Power supply, V/kW 220/0.8
Installation dimensions, WxDxH mm
513x445x590
(920 is the maximum height with a fully loaded polymer chamber and a full stack of loads with a handle for their installation)
Installation weight, not more than 70 kg
Produced in: Saint Petersburg