Nitrogen generator GA-200
Nitrogen generators are designed to create an inert atmosphere, fill low-pressure tanks, "rinse" test tubes with inert gas and other applications that require a large, by laboratory standards, nitrogen consumption with reduced requirements for its purity.
Technical specifications
Volume fraction of nitrogen is not less than % by volume (including impurities of inert gases — argon, neon, helium) 99.6
Volume fraction of oxygen is not more than, % vol. 0,4
Volume fraction of water vapor is not more than, % vol. 0.002 (20 ppm)
Maximum nitrogen capacity is at least 200 (3.33 l/min), l/h
Maximum nitrogen outlet pressure is not less than 3.5 ati
The time of setting the operating mode for single-shift operation is no more than 30 min
Nominal inlet air pressure is 6 ati
Maximum allowable inlet air pressure is 8 ati
Minimum inlet air pressure is 5 ati
Compressed air consumption, at nominal inlet pressure and maximum flow rate is no more than, l/h, 1200 (20 l/min)
Power consumption is not more than, VA, 100
Overall dimensions (width x depth x height) are not more than, mm, 230x525x510;
Weight is not more than 26 kg
Working conditions:
ambient temperature, C°, from +10 to +35
Electric power supply from
single-phase AC mains voltage, V, 220± 10%
and frequency, Hz, 50 ± 1
Average full service life is at least 6 years
The generator for electrical safety meets the requirements of Class 1, type H according to GOST 12.2.025-76
Produced in: Moscow