Back to search results

Picoammeter A2-4

System meter of small currents. The device allows to make measurements in almost the entire range of currents. The picoammeter is characterized by high performance, which is important for its use as part of automated measuring systems.

Technical specifications:

Current measuring range 1x10-14 - 1x10-2 A
Error for the range 10-11, 10-10 ±[0.5+0.025(lk/lx-1)]%
for the range 10-9, 10-8 ±[0.25+0.01(lk/lx-1)]%
for the range 10-7, 10-6,10-5, 10-4,10-3, 10-2 ±[0,1+0,01( lk/lx-1)]%
The time of setting the readings is 2x10-2 - 2x10-5 s
Supply voltage 230 (±20) V, frequency 50 (± 0.5) Hz
Power consumption 10 VA
Overall dimensions (HxWxD) 105x284x366 mm
Weight 3.5 kg
Operating temperature 5 - 40 °C
Humidity 80% at 25 °C