Back to search results

The FLUORAT®-411 alarm is heated, with a built-in sample preparation system

Back to search results
The FLUORAT®-411 alarm is heated, with a built-in sample preparation system
Manufacturer:
Lumeks
Location:
Saint Petersburg
Other:
Technical specifications:
The power consumption of the alarm is not more than 150 watts
The output electrical signal goes directly from the cable entry (B6), according to the standard scheme (4-20ma and (or) RS 485)
Sample temperature is not more than 200 degrees C
Sample pressure is not more than 1.5 Atm
The pressure of the spent sample is not more than 0.5 Atm
Cooling water temperature 5-30 Deg C
Dimensions of the alarm (mm) 1200*600*400
The boundaries of the spectral range of optical radiation used for analysis, nm
240 - 380 Analysis time, from 1
Power supply - alternating voltage (220 ± 10%) V, (50± 2.5) Hz operating mode settings, min no more than 30
Display range , mg/l 0.01-30
The limit of the basic error for the trigger threshold (%) 20
The set threshold for triggering the alarm
0.1 – 30.9 in increments of 0.1 mg/l the probability of trouble-free operation during 5000 hours is not less than 0.95
average recovery time, no more than 2
hours average service life, at least 5 years
Suppliers have this product:
Люмэкс
Price